topics- functions - index - search

Pulse plethysmography

What Does This Channel Measure?

For analysis of pulse plethysmography signals the ECG needs to have been analyzed. ANSLAB extracts five signals from the raw data: SV (systolic volume, in volts), DV (diastolic volume, in volts), PA (pulse amplitude=SV-DV), PTT (pulse transit time, in msec, measured from the ECG R-wave to the steepest upstroke in the pulse signal) and HR (heart rate, in bpm). For editing purposes you mainly need to be concerned with PA, an index of sympathetically mediated vascular constriction (resulting in smaller amplitudes) and PTT, a mostly sympathetic index that is related to factors such as systolic arterial pressure, vascular compliance, and stroke volume. Shorter PTT values correspond to higher sympathetic activation of the cardiovascular system.
ANSLAB displays the five signals in separate boxes in one window. PA and PTT are displayed in the third and fourth boxes from the top.








By clicking on a graph, you can zoom in on one variable and draw exclusion boxes to exlude outliers:







 
What Kind of Artifacts are Common in this Channel?

Artifacts in PA (see example below) or PTT appear as unusually large spikes in either direction.  While ‘normal’ values of PA vary by subject depending on the amplifier gain settings, PTT values typically fall within a range from 200 to 350 msec. Spikes beyond this range are almost always artifacts. A variability of 100 msec or less within a subject due to heart rate variability is normal and seldomly requires serious editing.
When you are not sure if something is an artifact, look at the raw data by clicking in the lower left corner and select option [10]. You will then be asked to mark suspicious segments by clicking on the left and right border of the suspicious segment (which can be repeated several times).
By pressing <0> the segements will be displayed as raw data in EXAM for further inspection. The figure contains the raw signal (yellow) and blue lines depicting the occurrence of the R-wave in the ECG, the following minimum corresponding to the diastolic value of the arterial pressure wave and the following maximum corresponding to the systolic value of the arterial pressure wave. This display mode can help you identify, e.g., artifacts due to ceiling effects, which can happen if the amplifier gains were set to high for a subject.





How Are Artifacts Removed?

Editing PA and PTT is simple.  Click on the box display for the parameter you want to edit.  A new window displaying this data pops up.  Mark the outliers by clicking on an upper threshold left of the spike, hold the mouse button, and drag the cross to a lower threshold right of the spike. By clicking, you create a box that defines the left and right margins, within which all data above and below the box is set to missing value. If you accidentally exclude too much, click once in the lower left corner of the window.  Data will be restored and you can try again.  By comparing the new mean to the original mean (displayed with the y-axis label), you can decide whether the data you excluded were outliers changing the mean.  If not, no further editing is necessary.  Click in the lower right corner to exit editing. Save reduced data.

What Qualities Must Be Preserved In Editing?

You only need to take out obvious outliers that would distort the mean, which is displayed at the left side of each graph. When in doubt, don’t exclude.  Also, don’t necessarily exclude ceiling or floor effects, since this might further distort the mean (because true extreme values are not only cut by a few percent, but left out of the mean estimation altogether). Linear interpolation after excluding data can be done by clicking to the right side of the plots, but usually is not necessary.